Drug, food and device recalls and enforcement reports from firms based in Ohio, pulled from FDA's openFDA database. Use it to check suppliers, contract manufacturers and competitors, and to see what FDA is finding near you.
| Report date | Firm | Type | Class | Product and reason |
|---|---|---|---|---|
| 2026-11-18 | 3B Medical, Inc. Dublin | Device | Class I | Luna G3 BPAP System Model/Catalog Number: LG3800-30VT Product Description: The Luna G3 BPAP System is a Bi-level PAP (Bi-level Positive Airway Pressure) The configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure when Targeted VT is enabled. |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q 40mm Screw Cap Feed Set, Model Number VED-052 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q 1000mL Bag Feed Set (LL), Model Number VED-051 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q 1000mL Bag Feed Set (PVC), Model Number VED-054 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q ENPlus Spike Feed Set, Model Numer VED-053 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q 500mL Bag Feed Set (PVC), Model Number VED-050 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-18 | Vesco Medical Llc Westerville | Device | Class II | Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8 The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the |
| 2026-11-11 | American Health Packaging Columbus | Drug | Class II | Cyanocobalamin Injection, USP, 1,000 mcg/mL, Rx Only, 25 x 1mL -Multiple-Dose Vials per carton, Mfg. by: American Regent, Inc., For Bluepoint Laboratories. Presence of particulate matter:American Regent, Inc. notified BluePoint Pharmaceuticals Laboratories regarding potential glass particle contamination (foreign glass fragment) identified in the |
| 2026-11-11 | Hikma Pharmaceuticals USA INC. Columbus | Drug | Class II | Ciprofloxacin Tablets, USP, 750 mg, 50-count bottle, Rx Only, Manufactured by: Hikma Pharmaceuticals, P.O. Box 182400, Amman, 11118-Jordan, Distributed by CGMP Deviations: N-nitroso-ciprofloxacin results above the FDA recommended interim control limit. |
| 2026-11-11 | Hikma Pharmaceuticals USA INC. Columbus | Drug | Class II | Ciprofloxacin Tablets, USP, 500 mg, 100-count bottle, Rx Only, Manufactured by: Hikma Pharmaceuticals, P.O. Box 182400, Amman, 11118-Jordan, Distributed by CGMP Deviations: N-nitroso-ciprofloxacin results above the FDA recommended interim control limit. |
| 2026-10-28 | Hikma Pharmaceuticals USA INC. Columbus | Drug | Class II | Abirtega (abiraterone acetate) USP, 250 mg, 120-count bottles, Rx only, Rx only, Mfd for: Hikma Pharmaceuticals USA Inc., Berkley Heights, NJ 07922 USA by Failed Tablet/Capsule Specifications: Failed tablet specifications for weight uniformity. |
| 2026-10-21 | American Health Packaging Columbus | Drug | Class II | Metoclopramide Tablets, USP, 5 mg, 100 Tablets (10 x 10), Rx Only, Distributed by: American Health Packaging, Columbus, Ohio 43217. NDC Carton: 60687-620-01 Presence of Foreign Tablets/Capsules |
| 2026-10-21 | Ferno-Washington Inc Wilmington | Device | Class II | Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821 Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or |
| 2026-10-21 | Steris Corporation Mentor | Device | Class II | RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115 If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the |
| 2026-10-21 | Steris Corporation Mentor | Device | Class II | RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117 If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the |
| 2026-10-21 | Sarnova HC, Llc Dublin | Device | Class II | Curaplex Capnography EtCO2/O2 Oral/Nasal Cannula Multipack (10 pk), Item Number 670272-KIT Affected kits were assembled with pediatric nasal cannula instead of the intended adult nasal cannula. |
| 2026-10-21 | Steris Corporation Mentor | Device | Class II | RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116 If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the |
| 2026-10-14 | Saint Francis Apizza LLC Cincinnati | Food/cosmetic | Class II | Saint Francis Apizza Frozen Pepperoni Pizza, Net Weight: 17 oz. (482g), packaged individually in vacuum-sealed plastic, UPC: 6 66186 27762 1 Undeclared soy allergen: soybean oil and soy lecithin |
| 2026-10-14 | Saint Francis Apizza LLC Cincinnati | Food/cosmetic | Class II | Saint Francis Apizza Frozen Cheese Pizza, Net Weight: 17 oz. (482g), packaged individually in vacuum-sealed plastic, UPC: 6 66186 27759 1 Undeclared soy allergen: soybean oil and soy lecithin |
| 2026-10-14 | Berlin Seeds Millersburg | Food/cosmetic | Class I | Berlin Seeds Alfalfa Sprouting Seeds, packaged in 1/4 lb., 1/2 lb., 1 lb., and 5 lb. resealable bags and 50 lb. woven poly bags Product contains alfalfa seed recalled by the supplier due to potential E. coli and/or Salmonella contamination. |
| 2026-10-14 | Saint Francis Apizza LLC Cincinnati | Food/cosmetic | Class II | Saint Francis Apizza Frozen Supreme Pizza, Net Weight: 17 oz. (482g), packaged individually in vacuum-sealed plastic, UPC: 6 66186 27760 7 Undeclared soy allergen: soybean oil and soy lecithin |
| 2026-10-14 | Saint Francis Apizza LLC Cincinnati | Food/cosmetic | Class II | Saint Francis Apizza Frozen Mushroom & Onion Pizza, Net Weight: 17 oz. (482g), packaged individually in vacuum-sealed plastic, UPC: 6 66186 27764 5 Undeclared soy allergen: soybean oil and soy lecithin |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control A software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation |
| 2026-10-14 | Steris Corporation Mentor | Device | Class II | VERIFY Biological Steam Test Pack Model/Catalog Number: S3069 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the |
| 2026-10-07 | LMSI, LLC Montgomery | Food/cosmetic | Class II | Kofinas Garlic Mediterranean Infused Extra Virgin Olive Oil, packaged in 3.4 fl oz (100mL), 8.5 fl oz (250mL), and 17 fl oz (500mL) glass bottles, UPC Product contains garlic essential oil not approved for culinary use |
| 2026-10-07 | American Contract Systems Inc Tiffin | Device | Class II | Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ABDOMINAL HYSTERECTOMY - 205989 Model No. MHAH55AM Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a |
| 2026-10-07 | American Nitrile Operations Llc Grove City | Device | Class II | Brand Name: Performance Pro Product Name: Performance Pro Model/Catalog Number: 72-RM50601,72-RM50602, 72-RM50603,72-RM50604 Product Description: Blue Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or |
| 2026-10-07 | American Contract Systems Inc Tiffin | Device | Class II | Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays CARDIAC HEART BUNDLE NS Model No. LMCH99P OPEN HEART Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a |
| 2026-10-07 | American Nitrile Operations Llc Grove City | Device | Class II | Brand Name: Opportunity Flex Product Name: Opportunity Flex Model/Catalog Number: 72-50900, 72-50901, 72-50902, 72-50903, 72-50904 Product Description: Blue Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or |
| 2026-10-07 | American Contract Systems Inc Tiffin | Device | Class II | Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a |
| 2026-10-07 | American Nitrile Operations Llc Grove City | Device | Class II | Brand Name: Opportunity Pro Product Name: Opportunity Pro Model/Catalog Number: 72-50800, 72-50801,72-50802,72-50803,72-50804 Product Description: Blue Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or |
| 2026-09-30 | Midmark Corporation Versailles | Device | Class II | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used. |
| 2026-09-30 | Midmark Corporation Versailles | Device | Class II | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used. |
| 2026-09-30 | Midmark Corporation Versailles | Device | Class II | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used. |
| 2026-09-30 | Midmark Corporation Versailles | Device | Class II | Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used. |
| 2026-09-16 | Ferrero USA North Canton | Food/cosmetic | Class II | Fannie May & XO Marshmallow Pixie S'mores Collection, 4.47oz Undeclared milk and pecans |
| 2026-09-09 | Savor Seasonings LLC Batavia | Food/cosmetic | Class I | Organic Cheddar Type Seasoning, NET WT: 50 lbs. Product is packaged in bulk poly-lined paper bag. Savor Seasonings, LLC. 4292 Armstrong Boulevard Batavia, OH Product tested positive for undeclared milk allergen. |
| 2026-09-09 | American Health Packaging Columbus | Drug | Class III | Buprenorphine Sublingual Tablets (C-III), 2 mg, 30 Tablets (3x10), Rx only, Distributed by: American Health Packaging, Columbus, Ohio 43217, Carton NDC# Labeling: Label Mix-up |
| 2026-09-02 | Sirna and Sons Produce, Inc Ravenna | Food/cosmetic | Class I | Sirna & Sons Produce PEPPER JALAPENO BREAK 1LB, repackaged into 1 lb. plastic bags, Product Number: 55255, Packed by Sirna & Sons Produce 7176 State Route 88 Potential Salmonella Javiana contamination |
| 2026-09-02 | Hikma Pharmaceuticals USA INC. Columbus | Drug | Class II | Ramipril Capsules USP, 10 mg, packaged in a) 100 Capsules (NDC 76282-673-01) and b) 500 Capsules (NDC 76282-673-05) bottles, Rx only, Manufactured for: Exelan Failed Impurities/Degradation Specifications: OOS result obtained for a process impurity, dicyclohexylurea (DCU). |
| 2026-09-02 | American Contract Systems Inc Tiffin | Device | Class II | American Contract System Convenience Kit, D & C PACK, Model/Catalog Number: LLPD90H The Convenience Kits contain Medline's Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not |
| 2026-09-02 | Midmark Corporation Versailles | Device | Class II | Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18. A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation. |
| 2026-09-02 | Midmark Corporation Versailles | Device | Class II | Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation. |
| 2026-09-02 | American Contract Systems Inc Tiffin | Device | Class II | American Contract System Convenience Kit, CV B PACK, Model/Catalog Number: TNCB10AN The Convenience Kits contain Medline s Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not |
| 2026-09-02 | American Contract Systems Inc Tiffin | Device | Class II | American Contract System Convenience Kit, CV B PACK, Model/Catalog Number: TNCB10AM The Convenience Kits contain Medline s Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not |
| 2026-09-02 | American Contract Systems Inc Tiffin | Device | Class II | American Contract System Convenience Kits, CV B PACK, Model/Catalog Number: TNCB10AP The Convenience Kits contain Medline s Red Rubber Intermittent Catheters (8Fr-22 Fr) and Red Rubber Intermittent Catheter Trays. Medline Industries, LP has identified representative samples did not |
| 2026-08-26 | American Contract Systems Inc Tiffin | Device | Class II | American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential |
| 2026-08-26 | Gentherm Medical, LLC Blue Ash | Device | Class II | Blanketrol II Model 222S; Thermal regulating system. Overheating of the 115V CPU board may cause the device to shut down. |
| 2026-08-26 | American Contract Systems Inc Tiffin | Device | Class II | American Contract Systems medical procedure kits and trays labeled as: 1) CRANI PACK - 208400, Model Item Number UDCP57AI; 2) SPINE PACK, Model Item The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential |
| 2026-08-26 | Gentherm Medical, LLC Blue Ash | Device | Class II | Blanketrol III, Model 233; Thermal regulating system. Overheating of the 115V CPU board may cause the device to shut down. |
What the recall classes mean
- Class I: a reasonable probability that the product will cause serious health problems or death.
- Class II: may cause temporary or medically reversible health problems, or the probability of serious problems is remote.
- Class III: not likely to cause adverse health consequences, but violates FDA rules.
If you manufacture in Ohio
Recalls often trace back to the same CGMP gaps that show up on Form 483s: microbial control and water systems, component testing, investigations and labeling. The 483 citation library explains each one in plain language, and FDA inspection readiness is how Claraxis helps plants find gaps before an investigator does.