FDA recalls in Massachusetts: the last 120 days.

Drug, food and device recalls and enforcement reports from firms based in Massachusetts, pulled from FDA's openFDA database. Use it to check suppliers, contract manufacturers and competitors, and to see what FDA is finding near you.

54enforcement reports
15firms
7Class I (most serious)
12drug reports
Report dateFirmTypeClassProduct and reason
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Liver System for organ transport OCS Solution Delivery Cassettes Model/Catalog Number: 1467 Product Description: OCS Solution Delivery Cassettes
Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Liver System for organ transport OCS Liver Perfusion Set Model/Catalog Number: 3200 Product Description: The recalled product is a standard off the
Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has
2026-11-18Straumann USA LLC
Andover
DeviceClass IIStraumann BLC Implants, SLActive, RXD. Endosseous Dental Implant. WB Healing Abutment, O6, GH1.5, AH2,Ti. Article Numbers: 035.9208S, 035.9210S, 035.9214S
Potential exposure to contamination, possibly compromising the sterility and cleanliness.
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Liver System for organ transport OCS Liver Solution Infusion Set Model/Catalog Number: 3457 Product Description: OCS Liver Solution Infusion Set
Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Liver Perfusion Set Model/Catalog Number: 3200 Software Version: rev 3 OCS Liver Perfusion Module included in the Liver Perfusion Set used for organ
The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Heart System OCS Heart Instrumentation Tool Set Model/Catalog Number: 1401 OCS Heart Instrumentation Tool Set, accessories intended to be used
Incorrectly assembled device kits that include incorrect components.
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Liver System OCS Liver Perfusion Set Model/Catalog Number: 3200 OCS Liver Perfusion Sets containing OCS Liver Perfusion Initiation Set and OCS Liver
Incorrectly assembled device kits that include incorrect components.
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Liver Perfusion Set Model/Catalog Number: 3200 OCS Liver Perfusion Sets containing OCS Liver Instrumentation Tool Set and OCS Liver Perfusion
Incorrectly assembled device sets with incorrect accessories.
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Heart System OCS Heart Perfusion Set Model/Catalog Number: 1200 OCS Heart Perfusion Sets containing OCS Heart Cardioplegia Arrest Set are
Incorrectly assembled device kits that include incorrect components.
2026-11-18Straumann USA LLC
Andover
DeviceClass IIStraumann Healing Abutments, Ti, Endosseous Dental Implant Abutments. Article Numbers: 064.4202S, 064.4204S, 064.4214S, 064.8212S, 064.8213S, 064.8215S.
Potential exposure to contamination, possibly compromising the sterility and cleanliness.
2026-11-18Boston Scientific Corporation
Marlborough
DeviceClass IIAMS 700 CX/LGX Rear Tip Extenders (RTEs), Snapcone, CX/LGX, 0.5 cm. UPN: 72404320.
Potential for the Rear Tip Extender (RTE) size contained within the package to differ from the size identified on the product labeling. Certain packages labeled as containing a 0.5 cm RTE may contain
2026-11-18Hologic, Inc
Marlborough
DeviceClass IIHorizon QDR X-Ray Bone Densitometer; Models: 1. Horizon W, 2. Horizon-WI;
A nonconformance was identified for systems manufactured from December 2024 to January 2025 that were shipped without the validated VGA cables. Some installations were completed using VGA cables that
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Heart Perfusion Set Model/Catalog Number: 1200 Software Version: rev 3 Product Description: OCS Heart Perfusion Module, included in the Heart
The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying
2026-11-18TransMedics, Inc.
Andover
DeviceClass IIOCS Lung Mission Critical Plus Perfusion Set Model/Catalog Number: 2200 Software Version: rev 1 OCS Lung MC+ Perfusion Module included in the Lung MC+
The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying
2026-11-11Fresenius Medical Care Renal Therapies Group, LLC
Lawrence
DrugClass IIDELFLEX Peritoneal Dialysis Solution with 4.25% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,500 mL Single dose bag, Sterile and
Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.
2026-11-11Fresenius Medical Care Renal Therapies Group, LLC
Lawrence
DrugClass IIDELFLEX Peritoneal Dialysis Solution with 2.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay"safe Exchange Set, 2,500 mL Single dose bag, Sterile and
Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.
2026-11-11Fresenius Medical Care Renal Therapies Group, LLC
Lawrence
DrugClass IIDELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,000 mL - Single Dose Bag, Sterile
Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.
2026-11-11Fresenius Medical Care Renal Therapies Group, LLC
Lawrence
DrugClass IIDELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,500 mL Single dose bag, Sterile and
Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.
2026-11-11Fresenius Medical Care Renal Therapies Group, LLC
Lawrence
DrugClass IIDELFLEX Peritoneal Dialysis Solution with 2.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,000 mL Single dose bag, Sterile and
Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots.
2026-11-04Abiomed, Inc.
Danvers
DeviceClass IAutomated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.
Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to
2026-11-04Abiomed, Inc.
Danvers
DeviceClass IImpella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to
2026-11-04Abiomed, Inc.
Danvers
DeviceClass IImpella Connect Optical AIC, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to
2026-11-04Philips North America
Cambridge
DeviceClass IIModel Number: 712010.
Potential for the device to fail immediately after an exposure, causing the system to remain stationary.
2026-10-28Philips North America
Cambridge
DeviceClass IIPhilips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central
2026-10-21DEANO'S PASTA
Somerville
Food/cosmeticClass IDeano's Pasta; Pear and Pecorino Triangoli Ravioli, 1lb.; Ingredients: Dough: Semolina (Wheat, Niacin, Thiamine, Mononitrate, Riboflavin, Folic Acid), Water
Undeclared Lobster
2026-10-21Instrumentation Laboratory
Bedford
DeviceClass IHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
2026-10-21Magellan Diagnostics, Inc.
North Billerica
DeviceClass IIILeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the
2026-10-21Instrumentation Laboratory
Bedford
DeviceClass IHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
2026-10-21Boston Scientific Corporation
Marlborough
DeviceClass IINaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;
Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.
2026-10-14PHILIPS MEDICAL SYSTEMS
Cambridge
DeviceClass IISpectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
2026-10-14LeMaitre Vascular, Inc.
Burlington
DeviceClass IITufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
2026-10-07Philips North America
Cambridge
DeviceClass IIIngenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export
2026-10-07Philips North America
Cambridge
DeviceClass IIMR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America
Cambridge
DeviceClass IIIngenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export
2026-10-07Philips North America
Cambridge
DeviceClass IIEvolution upgrade 1.5T amd Evolution upgrade 3.0T; Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782166 (1.5T), 4. 782201(1.5T), 5. 782117
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America Llc
Cambridge
DeviceClass IIPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient
2026-10-07Philips North America
Cambridge
DeviceClass IIUpgrade dStream; Model number (REF): 782127; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America
Cambridge
DeviceClass IIMR 5300; Model number (REF): 1. 782110, 2. 782152, 3. 782156, 4. 782196; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America
Cambridge
DeviceClass IISmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America
Cambridge
DeviceClass IIIngenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export
2026-10-07Philips North America
Cambridge
DeviceClass IIIngenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7.
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export
2026-10-07Philips North America
Cambridge
DeviceClass IIIngenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7.
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America
Cambridge
DeviceClass IIUpgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-10-07Philips North America
Cambridge
DeviceClass IISmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X; Model number (REF): 1.
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is
2026-09-30Philips North America Llc
Cambridge
DeviceClass IIAchieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system
2026-09-30Abiomed, Inc.
Danvers
DeviceClass IIAutomated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal
2026-09-23Safecor Health, LLC
Woburn
DrugClass IIFluphenazine HCl Oral Solution (Concentrate), USP, 15 mg per 3 mL, Delivers: 3 mL, PAI, Alcohol 14% by volume, For Oral Use Only, Rx Only, Pkg by: Safecor
Failed Stability Specifications
2026-09-23Safecor Health, LLC
Woburn
DrugClass IIFluphenazine HCl Oral Solution (Concentrate) , USP, 20 mg per 4 mL, Delivers: 4 mL, Oral Elixir, PAI, Alcohol 14% by volume, For Oral Use Only, Rx Only, Pkg
Failed Stability Specifications
2026-09-23Safecor Health, LLC
Woburn
DrugClass IIFluphenazine HCl Oral Solution (Concentrate), USP, 10 mg per 2 mL, Delivers: 2 mL, Oral Elixir, PAI, Alcohol 14% by volume, For Oral Use Only, Rx Only, Pkg by
Failed Stability Specifications
2026-09-23Safecor Health, LLC
Woburn
DrugClass IIFluphenazine HCl Oral Elixir, USP, 5 mg per 10 mL, Delivers: 10 mL, PAI, Alcohol 14% by volume, For Oral Use Only, Rx Only, Pkg by: Safecor Health, Woburn, MA
Failed Stability Specifications
2026-09-23Safecor Health, LLC
Woburn
DrugClass IIFluphenazine HCl Oral Elixir, USP, 5 mg per 1 mL, Delivers: 1 mL, PAI, Alcohol 14% by volume, Raspberry Flavor, For Oral Use Only, Rx Only, Pkg by: Safecor
Failed Stability Specifications
2026-09-23Safecor Health, LLC
Woburn
DrugClass IIFluphenazine HCl Oral Elixir, USP, 2.5 mg per 5 mL, Delivers: 5 mL, PAI, Alcohol 14% by volume, Raspberry Flavor, For Oral Use Only, Rx Only, Pkg by: Safecor
Failed Stability Specifications
2026-09-02Safecor Health, LLC
Woburn
DrugClass IChildren's Motrin (Ibuprofen) Oral Suspension, 150 mg per 7.5 mL, Delivers: 7.5 mL, Conc: 20 mg/mL, Oral Syringes, Berry Flavor/Alcohol Free, Mfg By: MCNEIL
Product Mixup; syringe filled with Atovaquone instead of Ibuprofen
2026-08-12CHEER PACK
West Bridgewater
Food/cosmeticClass IIFlexible food pouch (3-layer foil, spouted, #CP6088), manufactured for packaging Pumpkin Tree Peter Rabbit Organics Banana & Strawberry, 4 oz food puree, UPC
Flexible food pouch may release thin plastic strips into food.

What the recall classes mean

  • Class I: a reasonable probability that the product will cause serious health problems or death.
  • Class II: may cause temporary or medically reversible health problems, or the probability of serious problems is remote.
  • Class III: not likely to cause adverse health consequences, but violates FDA rules.

If you manufacture in Massachusetts

Recalls often trace back to the same CGMP gaps that show up on Form 483s: microbial control and water systems, component testing, investigations and labeling. The 483 citation library explains each one in plain language, and FDA inspection readiness is how Claraxis helps plants find gaps before an investigator does.